Reference Manual
00809-0600-4026, Rev BA
December 2011
4-7
Rosemount Process Radar
Figure 4-6. Surface threshold is
too high
Figure 4-7. Surface threshold is
too low
Recommended Actions
Adjust ATC threshold to detect the actual surface peak.
Alternatively adjust Upper Null Zone.
• Interface pulse not detected
The Surface Threshold is
above the Surface peak
Surface Threshold=
Amplitude Threshold
Curve (ATC)
Reference Threshold
Amplitude, mV
Distance, m
Actual surface
Disturbing
echo
misinterpreted
as product
surface
Amplitude, mV
Distance, m
Reference Threshold
Surface Threshold=
Amplitude Threshold
Curve (ATC)