32 Transmitter Configuration and Use: Series 1000 and 2000 Transmitters
Flowmeter Startup continued
Figure 5-2 Display – Loop test procedure
Set MAO1
Set MAO2
OFF-LINE MAINT
Scroll and Select
simultaneously for 4 seconds
Set FO
Set DO1
Set DO2
1 KHz
10 KHz
Read output at
receiving device
Read output at
receiving device
Loop test successful
Select to stop simulation
Scroll
Exit
Check output wiring
Troubleshoot output
Correct?
Read output at
receiving device
ON
OFF
4 mA
12 mA
20 mA
Ye s No output
Correct?
Correct?
OFF-LINE SIM
Scroll
Scroll
Select
Select
Select
Scroll
Scroll
Select
Select
Scroll
Select
Scroll
Select
Select
Vue de la page 41