Transmitter Configuration and Use: Series 1000 and 2000 Transmitters 121
Troubleshooting continued
Troubleshooting DiagramsDefaultsCompensation
A112 Series 1000/2000
software upgrade
recommended
NA Contact Micro Motion to get a Series 1000/2000
transmitter software upgrade. See Section 10.3.
Note that the device is still functional.
A113 Analog output 2
saturated
Analog 2 Saturated See Section 10.17.
A114 Analog output 2 fixed Analog 2 Fixed Exit mA output loop test. See Section 5.3.
Exit mA output trim. See Section 5.4.
Check to see if the output has been fixed via digital
communication.
A115 External input error External Input Error HART polling connection to external device has
failed. Ensure that external device is available:
• Verify device operation.
• Verify wiring.
Verify polling configuration. See Section 9.4.
A116 API temperature out of
limits
API temperature out of
limits
Verify process.
Verify API reference table and temperature
configuration. See Section 8.5.
A117 API density out of limits API density out of limits Verify process.
Verify API reference table and density configuration.
See Section 8.5.
A118 Discrete output 1 fixed DO1 Fixed Exit discrete output loop test. See Section 5.3.
A119 Discrete output 2 fixed DO2 Fixed Exit discrete output loop test. See Section 5.3.
A120 Bad fit data fault. Bad fit data fault Verify enhanced density configuration.
A121 Extrapolation warning Extrapolation warning Verify process temperature.
Verify process density.
Verify enhanced density configuration.
NA Density FD cal in
progress
NA Be advised that density calibration is in progress.
NA Density 1st point cal in
progress
NA Be advised that density calibration is in progress.
NA Density 2nd point cal in
progress
NA Be advised that density calibration is in progress.
NA Density 3rd point cal in
progress
NA Be advised that density calibration is in progress.
NA Density 4th point cal in
progress
NA Be advised that density calibration is in progress.
NA Mech. zero cal in
progress
NA Be advised that zero calibration is in progress.
NA Flow is in reverse
direction
NA Be advised that the process is flowing in reverse
direction.
Table 10-5 Status alarms and remedies (continued)
Alarm
code Communicator ProLink II software Possible remedy
Commentaires sur ces manuels