172 Transmitter Configuration and Use: Series 1000 and 2000 Transmitters
Configuring the Enhanced Density Application continued
D.7 Modifying a density curve
An existing density curve may be modified. The following parameters can be modified without
affecting the enhanced density calculations:
Curve name
Concentration unit label and optional text string
Extrapolation alarm limit
Note: As the value for extrapolation alarm limit is increased, the probability of inaccurate enhanced
density calculations also increases. Micro Motion recommends using the default value for
extrapolation alarm limit.
Do not change any other parameters. In particular, if you change the derived variable, all data are
erased for all existing curves.
D.8 Saving a density curve
Micro Motion recommends that all curves be saved to a file.
To save a curve to a file:
1. In the
ProLink/Configuration window, click on the ED Setup tab.
2. Use the
Curve being configured dropdown list to specify the curve to save, and click Apply.
3. Click the
Save this curve to a file button and specify the file name and location.
4. Repeat these steps for all density curves on your transmitter.