Transmitter Configuration and Use: Series 1000 and 2000 Transmitters 161
Calibrating the Transmitter continued
Enhanced Density Model 2500 CIOModel 1500 ANCalibration
Figure C-2 D1 and D2 density calibration – Communicator
Close shutoff valve
downstream from sensor
Fill sensor with D1 fluid
Diag/Service
Calibration
Density cal
Dens Pt 1
Enter density of D1 fluid
Perform Cal
Fill sensor with D2 fluid
HOME
Calibration in Progress
message
Density Calibration
Complete
message
On-Line Menu
OK
OK
Diag/Service
Calibration
Density cal
Dens Pt 2
Enter density of D2 fluid
Perform Cal
HOME
Calibration in Progress
message
Density Calibration
Complete
message
OK
OK