160 Transmitter Configuration and Use: Series 1000 and 2000 Transmitters
Calibrating the Transmitter continued
Figure C-1 D1 and D2 density calibration – ProLink II
Fill sensor with D1 fluid
Calibration
Density
Density cal - Point 1
Do Cal
Enter density of D1 fluid
Fill sensor with D2 fluid
Calibration in Progress
light turns red
ProLink
Calibration in Progress
light turns green
Calibration
Density
Density cal - Point 2
Do Cal
Enter density of D2 fluid
Calibration in Progress
light turns red
ProLink
Calibration in Progress
light turns green
Close shutoff valve
downstream from sensor
Close Close