Transmitter Configuration and Use: Series 1000 and 2000 Transmitters 171
Configuring the Enhanced Density Application continued
Enhanced Density Model 2500 CIOModel 1500 ANCalibration
D.6 Density curve trim
Density curve trim is used to bring transmitter output closer to reference values. Density curve trim
can be used for standard curves, custom curves, and user-defined curves. Density curve trim is an
optional procedure.
To trim a density curve:
1. In the
ED Setup window, click on the Show Advanced User Options button. The window
shown in Figure D-3 is displayed.
Figure D-3 ED Setup window – Advanced user options
2. Select the curve to be trimmed.
3. For the configured derived variable (mass concentration, density at reference temperature,
etc.), compare transmitter data with laboratory samples at a few temperature and concentration
points.
4. Calculate the slope and offset required to bring the transmitter output closer to the laboratory
values.
5. Enter the slope and offset values in the
Trim Slope and Trim Offset textboxes.
6. Click
Apply.
Slope and offset
textboxes