CAUTION!
Do not lift the sensor by the electronics or purge connections. Lifting the sensor by the
electronics or purge connections can damage the device.
2.2 Attach extended electronics
If you ordered a sensor with extended electronics, you will need to install the extender
onto the sensor case.
Note
Extended core processors are matched at the factory to specific sensors. Keep each core processor
together with the sensor with which it was shipped.
CAUTION!
Keep the extender and feedthrough clean and dry. Moisture or debris in the extender or
feedthrough can damage electronics and result in measurement error or flowmeter failure.
Procedure
1. Remove and recycle the plastic cap from the feedthrough on the sensor. See
Figure 2-2.
Mounting
Installation Manual 5