Page 1 - Density Sensors

Installation Manual20002298, Rev CCDecember 2010Micro Motion® F-Series Coriolis Flow andDensity Sensors

Page 2

Feedthrough and extender componentsFigure 2-2: ABCDEFGHA. Transmitter or core processorB. ExtenderC. Plastic plugD. Plastic capE. O-ringF. Clamping

Page 3 - Contents

3 WiringTopics covered in this chapter:•Options for wiring•Connect 4-wire cable•Connect 9-wire cable3.1 Options for wiringThe wiring procedure you fol

Page 4

Step 1: Cable preparationCable layoutRun conduit to sensorMetal conduitWrap the drain wires twice around the shield and cut off the excess drain wires

Page 5 - 1 Planning

Step 2: Shield terminationAssemble the Gland1. Fold the shield or braid back over the clamping insert and 1/8 inch (3 mm) past the O-ring.2. Install t

Page 6 - Environmental limits

Step 3: Terminal connectionsConnect the wires to the transmitter terminals(see the transmitter manual)Reinstall and tighten the core processor coverCo

Page 7 - Installation Manual 3

3. Match the wires color for color. For wiring at the transmitter or remote coreprocessor, refer to the transmitter documentation.4. Tighten the screw

Page 8 - 2 Mounting

4 GroundingThe sensor must be grounded according to the standards that are applicable at the site.The customer is responsible for knowing and complyin

Page 9 - Mounting

5 Supplementary informationTopics covered in this chapter:•Purge the sensor case•Pressure ratings for EN-1092 flanges5.1 Purge the sensor caseIf the s

Page 10

4. Connect the supply of nitrogen or argon gas to the inlet purge connection or openinlet purge line. Leave the outlet connection open.• Exercise caut

Page 11 - 3 Wiring

Pressure ratings for EN-1092 flangesFigure 5-1: 120100806040200AB and CDE0 50 150 200 250 300 350 400 427°Cbar160A. EN 1092-1 PN40 Forms B1 & D;

Page 12 - 8 Micro Motion F-Series

Safety and approval informationThis Micro Motion product complies with all applicable European directives when properly installed in accordance with t

Page 13 - Installation Manual 9

*20002298*20002298Rev CC2010Micro Motion Inc. USAWorldwide Headquarters7070 Winchester CircleBoulder, Colorado 80301T +1 303-527-5200T +1 800-522-6277

Page 14 - 3.3 Connect 9-wire cable

ContentsChapter 1 Planning ...

Page 15

Contentsii Micro Motion F-Series

Page 16 - 4 Grounding

1 PlanningTopics covered in this chapter:•Installation checklist•Best practices•Environmental limits1.1 Installation checklist□Make sure that the haza

Page 17 - 5 Supplementary information

□For best performance, follow Micro Motion recommendations for sensor orientation(see Table 1-3). The sensor will work in any orientation as long as t

Page 18 - Supplementary information

Environmental limits for F-Series sensorsFigure 1-1: Maximum process temperature in °F (°C)176 (80)140 (60)104 (40)68 (20)32 (0)–4 (–20)–40 (–40)–76

Page 19 - Installation Manual 15

2 MountingTopics covered in this chapter:•Mount the sensor•Attach extended electronics2.1 Mount the sensorUse your common practices to minimize torque

Page 20 - *20002298*

CAUTION!Do not lift the sensor by the electronics or purge connections. Lifting the sensor by theelectronics or purge connections can damage the devic