Reference Manual
00809-0200-4801, Rev CA
July 2010
B-11
Rosemount 3051S
Minimum Span Limits
Transmitter with Coplanar Sensor Module (Single Variable)
Transmitter with In-Line
Sensor Module
Transmitter with MultiVariable Sensor Module
(3051SMV_ _1 or 2, 3051SF_1, 2, 5, or 6)
Process Temperature RTD Interface
(3051SMV_ _1 or 3, 3051SF_1, 3, 5 or 7)
(1)
Lower (LRL) Upper (URL)
-328 °F (-200 °C) 1562 °F (850 °C)
(1) Transmitter is compatible with any Pt 100 RTD sensor. Examples of compatible RTDs include Rosemount Series 68 and 78 RTD Temperature Sensors.
Range
DP Sensor
(1)
(3051S_CD, 3051SMV_ _3 or 4,
3051SF_D, 3, 4 or 7, 3051S_LD)
(1) 3051SF flowmeters only available with ranges 1, 2, and 3.
GP Sensor
(3051S_CG, 3051S_LG)
AP Sensor
(3051S_CA, 3051S_LA)
Ultra &
Ultra for Flow
Classic Ultra Classic Ultra Classic
0 0.1 inH
2
O
(0,25 mbar)
0.1 inH
2
O
(0,25 mbar)
N/A N/A 0.167 psia
(11,5 mbar)
0.167 psia
(11,5 mbar)
1 0.5 inH
2
O
(1,24 mbar)
0.5 inH
2
O
(1,24 mbar)
0.5 inH
2
O
(1,24 mbar)
0.5 inH
2
O
(1,24 mbar)
0.3 psia
(20,7 mbar)
0.3 psia
(20,7 mbar)
2 1.3 inH
2
O
(3,11 mbar)
2.5 inH
2
O
(6,23 mbar)
1.3 inH
2
O
(3,11 mbar)
2.5 inH
2
O
(6,23 mbar)
0.75 psia
(51,7 mbar)
1.5 psia
(103,4 mbar)
3 5.0 inH
2
O
(12,4 mbar)
10.0 inH
2
O
(24,9 mbar)
5.0 inH
2
O
(12,4 mbar)
10.0 inH
2
O
(24,9 mbar)
4 psia
(275,8 mbar)
8 psia
(0,55 bar)
4 1.5 psi
(103,4 mbar)
3.0 psi
(206,8 mbar)
1.5 psig
(103,4 mbar)
3.0 psig
(206,8 mbar)
20 psia
(275,8 mbar)
40 psia
(2,76 bar)
5 10.0 psi
(689,5 mbar)
20.0 psi
(1,38 bar)
10.0 psig
(689,5 mbar)
20.0 psig
(1,38 bar)
N/A N/A
Range
GP Sensor
(3051S_TG)
AP Sensor
(3051S_TA)
Ultra Classic Ultra Classic
1 0.3 psig (20,7 mbar) 0.3 psig (20,7 mbar) 0.3 psia (20,7 mbar) 0.3 psia (20,7 mbar)
2 0.75 psig (51,7 mbar) 1.5 psig (0,103 bar) 0.75 psia (51,7 mbar) 1.5 psia (0,103 bar)
3 4 psig (275,8 mbar) 8 psig (0,55 bar) 4 psia (275,8 mbar) 8 psia (0,55 bar)
4 20 psig (1,58 bar) 40 psig (2,76 bar) 20 psia (1,58 bar) 40 psia (2,76 bar)
5 1000 psig (68,9 bar) 2000 psig (137,9 bar) 1000 psia (68,9 bar) 2000 psia (137,9 bar)
Range
DP Sensor
Ultra for Flow Classic MV
1 0.5 inH
2
O (1,24 mbar) 0.5 inH
2
O (1,24 mbar)
2 1.3 inH
2
O (3,11 mbar) 2.5 inH
2
O (6,23 mbar)
3 5.0 inH
2
O (12,4 mbar) 10.0 inH
2
O (24,9 mbar)
Range
Static Pressure Sensor (GP/AP)
Ultra for Flow Classic MV
3 4.0 psi (276 mbar) 8.0 psi (522 mbar)
4 18.13 psi (1,25 bar) 36.26 psi (2,50 bar)