Emerson Network Router Manuel d'utilisateur Page 58

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 96
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 57
52 ProLink
®
II Software for Micro Motion
®
Transmitters
Transmitter Configuration, Characterization, and Calibration
Figure 5-5 D1 and D2 density calibration
Fill sensor with D1 fluid
Calibration
Density cal - Point 1
Do Cal
Enter density of D1 fluid
Fill sensor with D2 fluid
Calibration in Progress
light turns red
ProLink
Calibration in Progress
light turns green
Calibration
Density cal - Point 2
Do Cal
Enter density of D2 fluid
Calibration in Progress
light turns red
ProLink
Calibration in Progress
light turns green
Close shutoff valve
downstream from sensor
Close Close
Vue de la page 57
1 2 ... 53 54 55 56 57 58 59 60 61 62 63 ... 95 96