18 ProLink
®
II Software for Micro Motion
®
Transmitters
Installation and Setup
Figure 2-12 Modbus/RS-485 connection to RS-485 terminals on MVD Direct Connect I.S. barrier
2.6 Configure ProLink II connection parameters
To connect to the transmitter, ProLink II must use connection parameters appropriate to the
transmitter.
If you are connecting to an MVD Direct Connect system, ProLink II can use any of the
supported communication settings listed in Table 2-2. The core processor auto-detects
incoming communications parameters and switches to match.
If you are connecting to a Model 2400S transmitter using the service port:
- For point-to-point connections, you can use a service port connection type.
- For multidrop network connections, you can use any Modbus/RS-485 connection type and
specify the transmitter’s Modbus address. The transmitter auto-detects incoming
communications parameters and switches to match. The service port auto-detection limits
are described in Table 2-3.
If you use a service port or HART/Bell 202 connection type, the connection parameters are
standard: when one of these connection types is specified, ProLink II automatically uses the
appropriate parameters.
RS-485 to RS-232
signal converter
Attach serial port or USB
adapter if necessary
RS-485/B
RS-485/A
I.S. barrier
Non-I.S. terminals
Notes:
Ensure that no other Modbus master devices are currently active on the network.
Before using ProLink II to communicate with the core processor, disconnect any wiring to a remote PLC.