Installation and Use Manual 47
Transmitter Configuration, Characterization, and Calibration
Meter Verification Transmitter TerminalsData LoggerTransmitter Configuration
5.4 Characterizing the meter
Characterizing the meter adjusts the transmitter to compensate for the unique traits of the sensor it is
paired with. The characterization parameters, or calibration parameters, describe the sensor’s
sensitivity to flow, density, and temperature.
5.4.1 When to characterize
If the transmitter and the sensor were ordered together as a Coriolis meter, then the meter has already
been characterized. You need to characterize the meter only if the transmitter and the sensor are being
paired together for the first time.
5.4.2 Characterization parameters
The characterization parameters that must be configured depend on your meter’s sensor type:
“T-Series” or “Other” (also referred to as “Straight Tube” and “Curved Tube,” respectively), as listed
in Table 5-1. The “Other” category includes all Micro Motion sensors except T-Series. Table 5-1 also
shows the location of each parameter within ProLink II.
The characterization parameters are provided on the sensor tag. The format of the sensor tag varies
depending on your sensor’s date of purchase. See Figures 5-2 and 5-3 for illustrations of newer and
older sensor tags.
Table 5-1 Sensor calibration parameters
Parameter ProLink II location
Sensor type
T- Se ri es O t h e r
K1 Configuration/Density ✓✓
(1)
(1) See the section entitled “Density calibration factors.
K2 Configuration/Density ✓✓
(1)
FD Configuration/Density ✓✓
(1)
D1 Configuration/Density ✓✓
(1)
D2 Configuration/Density ✓✓
(1)
Temp coeff (DT)
(2)
(2) On some sensor tags, shown as TC.
Configuration/Density ✓✓
(1)
Flowcal Configuration/Flow
(3)
(3) See the section entitled “Flow calibration values.
FCF and FT Configuration/Flow
(4)
(4) Older T-Series sensors. See the section entitled “Flow calibration values.
FCF Configuration/Flow
(5)
(5) Newer T-Series sensors. See the section entitled “Flow calibration values.
FTG Configuration/T-Series Config
FFQ Configuration/T-Series Config
DTG Configuration/T-Series Config
DFQ1 Configuration/T-Series Config
DFQ2 Configuration/T-Series Config