LIQ_MAN_ABR_5081T-HT/Rev. L January 2015
+(&@H6C)FAA=JFCC6?E"@@A-:C:?87@C**
-($*(%+ +$*%$%.
The sensor can be wired to the analyzer through a remote
junction box (PN 23550-00). Wire the extension cable and
sensor cable point-to-point. Refer to the sensor instruc-
tion manual for more details.
Factory-terminated (PN 23294-05) and unterminated (PN
9200276) connecting cable are available. The use of fac-
tory-terminated cable is strongly recommended. To pre-
pare unterminated cable for use, follow the instructions in
the sensor instruction manual.
For maximum EMI/RFI protection, the outer braid of the
sensor cable should be connected to the outer braided
shield of the extension cable. At the instrument, connect
the outer braid of the extension cable to earth ground.
&%-(-($
For general purpose areas, wire power as shown in Figure
3. For hazardous areas, please see hazardous area installa-
tion drawings.