3.7.3 Totalizer
(Fully Compensated Mass and Energy Flow Feature Board only):
The Totalizer tab is used to configure the settings associated with the Totalizer functionality
within the transmitter.
Figure 3-32. Variables - Totalizer tab
1. To turn the totalizer functionality on or off, select Start or Stop from the Mode drop
down menu under the Totalizer Setup heading. The totalizer Units may also be edited
under this heading.
2. Verify the Totalized Parameter and the Totalizer Maximum value. To edit theTotalized
Parameter, click on the Configure Totalizer button under the Totalizer Control heading.
Note
When the totalizer reaches its maximum value, it automatically resets to zero and continues
totalizing. The default maximum is a value equivalent to 4.29 billion pounds, actual cubic feet,
or BTU. To edit the Totalizer Maximum value, click on the Set Totalizer Maximum button under
the Totalizer Control heading.
3. To reset the Totalized Reading to zero, click the Reset Totalizer button under the Totalizer
Control heading.
4. Under the Custom Units Setup heading, the user may configure a custom unit for the
Totalized Reading. Custom units allow the totalizer rate to be displayed in units of
measure that are not standard in the 3051S MultiVariable.
Mass and Energy Flow Fast Keys 1, 4, 1, 3