Micro Motion
®
Model 2400S Transmitters 7
Digital communications
Host interface
All versions
Service port One service port for temporary connections (requires removing transmitter housing cover)
Uses RS-485 Modbus signal, 38.4 kilobaud, one stop bit, no parity
Address: 111 (not configurable)
Wireless If transmitter has display, service port can be accessed via serial port IrDA without removing
transmitter housing cover.
Model 2400S Analog
HART/Bell 202 HART signal is superimposed on the milliamp output (Channel A), and is available for host
system interface:
• Frequency: 1.2 and 2.2 kHz
• Amplitude: to 1.0 mA
• 1200 baud, one stop bit, odd parity
• Address: 0 (default), configurable
• Requires 250 to 600 Ω resistance
Model 2400S PROFIBUS-DP
PROFIBUS-DP Digital 2-way communication protocol
• Automatically recognizes network baud rate
• Address selectable by 3 rotary switches, or software selectable
Model 2400S DeviceNet
DeviceNet Digital 2-way communication protocol
• Address and baud rate selectable by 3 rotary switches (2 to select address, 1 to select
baud rate), or software selectable
Model 2400S Analog ProLink
®
II v2.9 (or later) from Micro Motion supports full device configuration.
HART DD file supports all functionality.
Model 2400S PROFIBUS-DP • ProLink II v2.9 (or later) from Micro Motion supports full device configuration.
• GSD file conforming to the PROFIBUS-DP specification
- Provides PROFIBUS Class 1 Master functions
- Enables reading and controlling all process data
• DD file conforming to PROFIBUS EDDL specification
- Provides PROFIBUS Class 2 Master functions
- Enables device configuration
- Supports Siemens Simatic PDM
Model 2400S DeviceNet • ProLink II v2.9 (or later) from Micro Motion supports full device configuration.
• EDS file conforming to the DeviceNet specification:
- Enables device configuration
Commentaires sur ces manuels