Emerson IFT9701 Manuel d'instructions Page 70

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 90
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 69
62 Model IFT9701 Transmitter Instruction Manual
Troubleshooting continued
If the transmitter indicates an over range or sensor failure condition, follow these steps:
1. Isolate the transmitter from devices that use transmitter outputs to control the flow loop.
2. Refer to Table 8-5 to diagnose the problem.
3. If troubleshooting fails to reveal why over range and/or sensor failure messages have appeared,
phone Micro Motion Customer Service. Phone numbers are listed on the title page of this
manual.
8.7 Slug flow
Programmed slug flow limits minimize inaccurate measurement caused by significant quantities
(slugs) of gas during liquid flow measurement or slugs of liquid during gas flow measurement. Such
conditions adversely affect sensor performance by causing erratic vibration of the flow tubes, which
in turn causes the transmitter to produce inaccurate flow signals.
Table 8-5 Troubleshooting over range and sensor failure conditions
Digital multimeter
(DMM) Cause(s) Corrective action(s)
Open or short from red to
brown at transmitter
Open or short from red to
brown at sensor
Faulty cable
Faulty drive coil in sensor
Moisture in sensor case or junction box
If open or short at transmitter terminals,
check cable
If open or short at sensor junction box,
return sensor to factory
Replace conduit and/or conduit seals
Repair cable
• Open or short from green to
white at transmitter
• Open or short from green to
white at sensor
Faulty cable
Faulty left pickoff in sensor
Moisture in sensor case or junction box
Open or short from blue to
gray at transmitter
Open or short from blue to
gray at sensor
Faulty cable
Faulty right pickoff in sensor
Moisture in sensor case or junction box
No open circuits
No short circuits
Transmitter cannot calculate offset of flow
signal
Ensure that the sensor is filled with fluid
Eliminate noise, then rezero
Completely shut off flow, then rezero
Transmitter cannot calculate flow rate Eliminate pipe stress, vibration, or
mechanical noise
Inappropriate density factor
Density < 0.200 g/cc or > 2.000 g/cc,
volume flow is being measured
Density > 2.000 g/cc, mass flow is being
measured
Erratic process density has caused flow
tubes to stop vibrating
Plugged flow tube
Transmitter cannot calculate density
Characterize density values for sensor
Monitor density
Bring density within sensor limit
Purge flow tubes
Eliminate pipe stress, vibration, or
mechanical noise
Open or short from yellow
to orange at transmitter
Open or short from yellow
to orange at sensor
Temperature outside sensor limit
Faulty cable
Faulty lead length compensator
Bring temperature within sensor limit
Monitor temperature
If open or short at transmitter terminals,
check cable
If open or short at sensor junction box,
return sensor to factory
Open or short from violet to
yellow at transmitter
Open or short from violet to
yellow at sensor
Faulty cable
Faulty RTD in sensor
Vue de la page 69
1 2 ... 65 66 67 68 69 70 71 72 73 74 75 ... 89 90

Commentaires sur ces manuels

Pas de commentaire