http://www.raihome.comInstruction ManualIM-106-340C, Rev 4.4July 2008Oxymitter 4000 Hazardous Area Oxygen Transmitter
Hazardous Area Oxymitter 4000iiInstruction ManualIM-106-340C, Rev. 4.4July 2008PREFACEThe purpose of this manual is to provide information concerning
8-10Instruction ManualIM-106-340C, Rev. 4.4July 2008Hazardous Area Oxymitter 4000Figure 8-6. Fault 4, A/D Comm Error Fault 4, A/D Comm ErrorMembrane K
Instruction Manual IM-106-340C, Rev. 4.4July 20088-11Hazardous Area Oxymitter 4000Figure 8-7. Fault 5, Open Heater Fault 5, Open HeaterFigure 8-7 show
8-12Instruction ManualIM-106-340C, Rev. 4.4July 2008Hazardous Area Oxymitter 4000Figure 8-8. Fault 6, High High Heater Temp Fault 6, High High Heater
Instruction Manual IM-106-340C, Rev. 4.4July 20088-13Hazardous Area Oxymitter 4000Figure 8-9. Fault 7, High Case TempFault 7, High Case TempFigure 8-9
8-14Instruction ManualIM-106-340C, Rev. 4.4July 2008Hazardous Area Oxymitter 4000Figure 8-10. Fault 8, Low Heater Temp Fault 8, Low Heater TempFigure
Instruction Manual IM-106-340C, Rev. 4.4July 20088-15Hazardous Area Oxymitter 4000Figure 8-11. Fault 9, High Heater Temp Fault 9, High Heater TempFigu
8-16Instruction ManualIM-106-340C, Rev. 4.4July 2008Hazardous Area Oxymitter 4000Figure 8-12. Fault 10, High Cell mV Fault 10, High Cell mVFigure 8-12
Instruction Manual IM-106-340C, Rev. 4.4July 20088-17Hazardous Area Oxymitter 4000Figure 8-13. Fault 11, Bad Cell Fault 11, Bad CellFigure 8-13 shows
8-18Instruction ManualIM-106-340C, Rev. 4.4July 2008Hazardous Area Oxymitter 4000Figure 8-14. Fault 12, EEprom Corrupt Fault 12, EEprom CorruptFigure
Instruction Manual IM-106-340C, Rev. 4.4July 20088-19Hazardous Area Oxymitter 4000Figure 8-15. Fault 13, Invalid Slope Fault 13, Invalid SlopeFigure 8
Instruction Manual IM-106-340C, Rev. 4.4July 2008Hazardous Area Oxymitter 4000http://www.raihome.comSection 1 Description and SpecificationsComponent
8-20Instruction ManualIM-106-340C, Rev. 4.4July 2008Hazardous Area Oxymitter 4000Figure 8-16. Fault 14, Invalid ConstantFault 14, Invalid ConstantFigu
Instruction Manual IM-106-340C, Rev. 4.4July 20088-21Hazardous Area Oxymitter 4000Figure 8-17. Fault 15, Last Calibration Failed Fault 15, Last Calibr
Hazardous Area Oxymitter 40008-22Instruction ManualIM-106-340C, Rev. 4.4July 2008HEATER NOT OPEN, BUT UNABLE TO REACH 736°C SETPOINTThe temperature se
Instruction Manual IM-106-340C, Rev. 4.4July 20088-23Hazardous Area Oxymitter 40002. The sensing cell is bolted to the end of the probe and uses a cor
Hazardous Area Oxymitter 40009-0Instruction ManualIM-106-340C, Rev. 4.4July 200838860078Calibration RecordForRosemount Analytical In-Situ O2ProbeProbe
Instruction Manual IM-106-340C, Rev. 4.4July 2008Hazardous Area Oxymitter 4000http://www.raihome.comSection 9 Maintenance and ServiceOverview . . .
Hazardous Area Oxymitter 40009-2Instruction ManualIM-106-340C, Rev. 4.4July 2008In applications with a heavy dust loading, the O2 probe diffusion elem
Instruction Manual IM-106-340C, Rev. 4.4July 20089-3Hazardous Area Oxymitter 4000Figure 9-1. Membrane Keypad• If using an IMPS 4000, enter a time inte
Hazardous Area Oxymitter 40009-4Instruction ManualIM-106-340C, Rev. 4.4July 2008• HART. Use the HART Communicator to access the O2 CALIBRATE menu and
Instruction Manual IM-106-340C, Rev. 4.4July 20089-5Hazardous Area Oxymitter 40003. If performing a manual calibration with the CALIBRATION RECOMMENDE
Hazardous Area Oxymitter 40001-2Instruction ManualIM-106-340C, Rev. 4.4July 2008Figure 1-1. Typical System Package 1. Instruction Manual2. IMPS 4000 I
Hazardous Area Oxymitter 40009-6Instruction ManualIM-106-340C, Rev. 4.4July 2008If the calibration was valid, the DIAGNOSTIC ALARMS LEDs will indicate
Instruction Manual IM-106-340C, Rev. 4.4July 20089-7Hazardous Area Oxymitter 40005. Remove the second calibration gas and cap off the calibration gas
Hazardous Area Oxymitter 40009-8Instruction ManualIM-106-340C, Rev. 4.4July 2008Removal and Replacement of Probe1. Remove.a. Turn off power to the sys
Instruction Manual IM-106-340C, Rev. 4.4July 20089-9Hazardous Area Oxymitter 40003. Disconnect the heater and signal wire connectors from the mating c
Hazardous Area Oxymitter 40009-10Instruction ManualIM-106-340C, Rev. 4.4July 2008Figure 9-3. Hazardous Area Oxymitter 4000 with Integral Electronics -
Instruction Manual IM-106-340C, Rev. 4.4July 20089-11Hazardous Area Oxymitter 4000Figure 9-4. Hazardous Area Oxymitter 4000 with Remote Electronics -
Hazardous Area Oxymitter 40009-12Instruction ManualIM-106-340C, Rev. 4.4July 20086. Follow the instructions in "Probe Installation" in Secti
Instruction Manual IM-106-340C, Rev. 4.4July 20089-13Hazardous Area Oxymitter 4000Figure 9-6. J8 Connector4. See Figure 9-6. Squeeze the sides of the
Hazardous Area Oxymitter 40009-14Instruction ManualIM-106-340C, Rev. 4.4July 2008Terminal Block ReplacementSee Figure 9-3 or Figure 9-4.1. Remove the
Instruction Manual IM-106-340C, Rev. 4.4July 20089-15Hazardous Area Oxymitter 4000Figure 9-7. Fuse Location10. Reconnect the J8 connector to the power
Instruction Manual IM-106-340C, Rev. 4.4July 20081-3Hazardous Area Oxymitter 4000System DescriptionThe Hazardous Area Oxymitter 4000 is designed to me
Hazardous Area Oxymitter 40009-16Instruction ManualIM-106-340C, Rev. 4.4July 20081. Follow the instructions in "Removal and Replacement of Probe&
Instruction Manual IM-106-340C, Rev. 4.4July 20089-17Hazardous Area Oxymitter 4000Figure 9-8. Heater Strut Assembly 8. See Figure 9-8. Grasp the wire
Hazardous Area Oxymitter 40009-18Instruction ManualIM-106-340C, Rev. 4.4July 2008Cell ReplacementThis paragraph covers O2 cell replacement. Do not att
Instruction Manual IM-106-340C, Rev. 4.4July 20089-19Hazardous Area Oxymitter 40001. Follow the instructions in "Removal and Replacement of Probe
Hazardous Area Oxymitter 40009-20Instruction ManualIM-106-340C, Rev. 4.4July 2008Figure 9-10. Ceramic Diffuser Element Replacement8. Apply a light coa
Instruction Manual IM-106-340C, Rev. 4.4July 20089-21Hazardous Area Oxymitter 4000Ceramic Diffusion Element ReplacementNOTEThis procedure applies to t
Hazardous Area Oxymitter 40009-22Instruction ManualIM-106-340C, Rev. 4.4July 20089. Thoroughly mix cement and insert tip of squeeze bottle into cement
Instruction Manual IM-106-340C, Rev. 4.4July 20089-23Hazardous Area Oxymitter 4000Contact and Thermocouple Assembly ReplacementSee Figure 9-11.1. Remo
Hazardous Area Oxymitter 40009-24Instruction ManualIM-106-340C, Rev. 4.4July 2008
Instruction Manual IM-106-340C, Rev. 4.4July 2008Hazardous Area Oxymitter 4000http://www.raihome.comSection 10 Replacement PartsProbe Replacement Part
Hazardous Area Oxymitter 40001-4Instruction ManualIM-106-340C, Rev. 4.4July 2008System ConfigurationHazardous Area Oxymitter 4000 units are available
Hazardous Area Oxymitter 400010-2Instruction ManualIM-106-340C, Rev. 4.4July 2008Table 10-1. Replacement Parts for Probe (Continued)* Includes pad an
Instruction Manual IM-106-340C, Rev. 4.4July 200810-3Hazardous Area Oxymitter 4000Table 10-1. Replacement Parts for Probe (Continued)Figure 10-2. Pro
Hazardous Area Oxymitter 400010-4Instruction ManualIM-106-340C, Rev. 4.4July 2008ELECTRONICS REPLACEMENT PARTSTable 10-2. Replacement Parts for Elect
Instruction Manual IM-106-340C, Rev. 4.4July 2008Hazardous Area Oxymitter 4000http://www.raihome.comSection 11 Optional AccessoriesHART Handheld 375
Hazardous Area Oxymitter 400011-2Instruction ManualIM-106-340C, Rev. 4.4July 2008ASSET MANAGEMENT SOLUTIONS (AMS)Asset Management Solutions (AMS) soft
Instruction Manual IM-106-340C, Rev. 4.4July 200811-3Hazardous Area Oxymitter 4000IMPS 4000 INTELLIGENT MULTIPROBE TEST GAS SEQUENCERFigure 11-3. IMPS
Hazardous Area Oxymitter 400011-4Instruction ManualIM-106-340C, Rev. 4.4July 2008SPS 4001B SINGLE PROBE AUTOCALIBRATION SEQUENCERFigure 11-4. SPS 4001
Instruction Manual IM-106-340C, Rev. 4.4July 200811-5Hazardous Area Oxymitter 4000O2 CALIBRATION GASFigure 11-5. Calibration Gas BottlesRosemount Anal
Hazardous Area Oxymitter 400011-6Instruction ManualIM-106-340C, Rev. 4.4July 2008
Instruction Manual IM-106-340C, Rev. 4.4July 2008Hazardous Area Oxymitter 4000http://www.raihome.comAppendix A Safety DataSafety Instructions . . . .
Instruction Manual IM-106-340C, Rev. 4.4July 20081-5Hazardous Area Oxymitter 4000For systems with one or two Hazardous Area Oxymitter 4000 units per c
Hazardous Area Oxymitter 4000A-2Instruction ManualIM-106-340C, Rev. 4.4July 2008SAFETY INSTRUCTIONS IMPORTANT SAFETY INSTRUCTIONS FOR THE WIRING AND I
Instruction Manual IM-106-340C, Rev. 4.4July 2008A-3Hazardous Area Oxymitter 4000DŮLEŽITÉ Bezpečnostní pokyny pro zapojení a instalaci zařízeníNásledu
Hazardous Area Oxymitter 4000A-4Instruction ManualIM-106-340C, Rev. 4.4July 2008VIGTIGTSikkerhedsinstruktion for tilslutning og installering af dette
Instruction Manual IM-106-340C, Rev. 4.4July 2008A-5Hazardous Area Oxymitter 4000BELANGRIJKVeiligheidsvoorschriften voor de aansluiting en installatie
Hazardous Area Oxymitter 4000A-6Instruction ManualIM-106-340C, Rev. 4.4July 2008BELANGRIJKVeiligheidsinstructies voor de bedrading en installatie van
Instruction Manual IM-106-340C, Rev. 4.4July 2008A-7Hazardous Area Oxymitter 4000WICHTIGSicherheitshinweise für den Anschluß und die Installation dies
Hazardous Area Oxymitter 4000A-8Instruction ManualIM-106-340C, Rev. 4.4July 2008ΣΗΜΑΝΤΙΚΟ Οδηγιεσ ασφαλειασ για την καλωδιωση και εγκατασταση τησ συσκ
Instruction Manual IM-106-340C, Rev. 4.4July 2008A-9Hazardous Area Oxymitter 4000OLULINE TEAVE Juhtmestiku ja seadme paigaldamisega seotud ohutusjuhis
Hazardous Area Oxymitter 4000A-10Instruction ManualIM-106-340C, Rev. 4.4July 2008TÄRKEÄÄTurvallisuusohje, jota on noudatettava tämän laitteen asentami
Instruction Manual IM-106-340C, Rev. 4.4July 2008A-11Hazardous Area Oxymitter 4000IMPORTANTConsignes de sécurité concernant le raccordement et l'
Hazardous Area Oxymitter 40001-6Instruction ManualIM-106-340C, Rev. 4.4July 2008Figure 1-4. Local Operator Interface (LOI) An optional Local Operator
Hazardous Area Oxymitter 4000A-12Instruction ManualIM-106-340C, Rev. 4.4July 2008FONTOS Biztonsági előírások a készülék vezetékeléséhez és üzembeállít
Instruction Manual IM-106-340C, Rev. 4.4July 2008A-13Hazardous Area Oxymitter 4000IMPORTANTENorme di sicurezza per il cablaggio e l'installazione
Hazardous Area Oxymitter 4000A-14Instruction ManualIM-106-340C, Rev. 4.4July 2008SVARBU šio prietaiso laidų prijungimo ir instaliacijos saugos instruk
Instruction Manual IM-106-340C, Rev. 4.4July 2008A-15Hazardous Area Oxymitter 4000SVARĪGI Drošības norādījumi šīs iekārtas pievienošanai un uzstādīšan
Hazardous Area Oxymitter 4000A-16Instruction ManualIM-106-340C, Rev. 4.4July 20081. Konnessjonijiet adegwati ta' l-ert għandhom isiru għall-pun
Instruction Manual IM-106-340C, Rev. 4.4July 2008A-17Hazardous Area Oxymitter 4000VIKTIGSikkerhetsinstruks for tilkobling og installasjon av dette uts
Hazardous Area Oxymitter 4000A-18Instruction ManualIM-106-340C, Rev. 4.4July 2008WAŻNE! Zalecenia dotyczące bezpieczeństwa w zakresie podłączania i in
Instruction Manual IM-106-340C, Rev. 4.4July 2008A-19Hazardous Area Oxymitter 4000IMPORTANTEInstruções de segurança para ligação e instalação deste ap
Hazardous Area Oxymitter 4000A-20Instruction ManualIM-106-340C, Rev. 4.4July 2008DÔLEŽITÉ Bezpečnostné pokyny pre zapojenie káblov a inštaláciu tohto
Instruction Manual IM-106-340C, Rev. 4.4July 2008A-21Hazardous Area Oxymitter 4000POMEMBNO Varnostna navodila za povezavo in vgradnjo napraveNaslednja
Instruction Manual IM-106-340C, Rev. 4.4July 20081-7Hazardous Area Oxymitter 4000Membrane KeypadThe membrane keypad, housed within the right side of t
Hazardous Area Oxymitter 4000A-22Instruction ManualIM-106-340C, Rev. 4.4July 2008IMPORTANTEInstrucciones de seguridad para el montaje y cableado de es
Instruction Manual IM-106-340C, Rev. 4.4July 2008A-23Hazardous Area Oxymitter 4000VIKTIGTSäkerhetsföreskrifter för kablage och installation av denna a
Hazardous Area Oxymitter 4000A-24Instruction ManualIM-106-340C, Rev. 4.4July 2008SAFETY DATA SHEET FOR CERAMIC FIBER PRODUCTSJULY 1, 1996SECTION I.
Instruction Manual IM-106-340C, Rev. 4.4July 2008A-25Hazardous Area Oxymitter 4000SECTION III. HAZARDOUS INGREDIENTSMATERIAL, QUANTITY, AND THRESHOL
Hazardous Area Oxymitter 4000A-26Instruction ManualIM-106-340C, Rev. 4.4July 2008EXPOSURE TO USED CERAMIC FIBER PRODUCTProduct which has been in servi
Instruction Manual IM-106-340C, Rev. 4.4July 2008A-27Hazardous Area Oxymitter 4000The International Agency for Research on Cancer (IARC) reviewed the
Hazardous Area Oxymitter 4000A-28Instruction ManualIM-106-340C, Rev. 4.4July 2008SECTION VIII. SPECIAL PROTECTION INFORMATIONRESPIRATORY PROTECTIONUse
Instruction Manual IM-106-340C, Rev. 4.4July 2008A-29Hazardous Area Oxymitter 4000IARC has recently reviewed the animal, human, and other relevant exp
Hazardous Area Oxymitter 4000A-30Instruction ManualIM-106-340C, Rev. 4.4July 2008GENERAL PRECAUTIONS FOR HANDLING AND STORING HIGH PRESSURE GAS CYLIND
Instruction Manual IM-106-340C, Rev. 4.4July 2008Hazardous Area Oxymitter 4000http://www.raihome.comAppendix B Return of MaterialRETURNING MATERIALIf
Hazardous Area Oxymitter 40001-8Instruction ManualIM-106-340C, Rev. 4.4July 2008Handling the OxymitterSystem ConsiderationsPrior to installing your Ha
Hazardous Area Oxymitter 4000B-2Instruction ManualIM-106-340C, Rev. 4.4July 2008
Instruction ManualIM-106-340C, Rev. 4.4July 2008Hazardous Area Oxymitter 4000http://www.raihome.comIndexAAbrasive Shield Assembly . . . 1-13Alarm Cont
Instruction ManualIM-106-340C, Rev. 4.4July 2008Hazardous Area Oxymitter 4000Index-2
WARRANTYRosemount Analytical warrants that the equipment manufactured and sold by it will, upon shipment, be free of defects in workmanship or materia
© 2008 Emerson Process Management. All rights reserved.Rosemount Analytical and the Rosemount Analytical logotype are registered trademarks of Rosemou
Instruction Manual IM-106-340C, Rev. 4.4July 20081-9Hazardous Area Oxymitter 4000Figure 1-6. Typical System Installation – Oxymitter 4000 with Integra
Hazardous Area Oxymitter 40001-10Instruction ManualIM-106-340C, Rev. 4.4July 2008Figure 1-7. Typical System Installation – Oxymitter 4000 with Remote
Instruction Manual IM-106-340C, Rev. 4.4July 20081-11Hazardous Area Oxymitter 4000A source of instrument air is optional at the Hazardous Area Oxymitt
Hazardous Area Oxymitter 40001-12Instruction ManualIM-106-340C, Rev. 4.4July 2008Figure 1-8. Model 751 Remote Powered Loop LCD Display PROBE OPTIONSDi
Instruction Manual IM-106-340C, Rev. 4.4July 20081-13Hazardous Area Oxymitter 4000Figure 1-10. Flame Arrestor Snubber Diffusion AssemblyFlame Arrestor
Hazardous Area Oxymitter 40001-14Instruction ManualIM-106-340C, Rev. 4.4July 2008Figure 1-11. Abrasive Shield AssemblyNOTEIn highly abrasive applicati
Instruction Manual IM-106-340C, Rev. 4.4July 20081-15Hazardous Area Oxymitter 4000SPECIFICATIONSFisher-Rosemount has satisfied all obligations coming
Hazardous Area Oxymitter 40001-16Instruction ManualIM-106-340C, Rev. 4.4July 2008SignalsAnalog Output/HART 4-20 mA isolated from power supply, 950 ohm
Instruction Manual IM-106-340C, Rev. 4.4July 20081-17Hazardous Area Oxymitter 4000Table 1-1. Product MatrixOXT4C Oxymitter 4000 Explosion Proof - In
Hazardous Area Oxymitter 40001-18Instruction ManualIM-106-340C, Rev. 4.4July 2008NOTES:(1)Recommended uses: High velocity particulates in flue stream,
Instruction Manual IM-106-340C, Rev. 4.4July 20081-19Hazardous Area Oxymitter 4000Table 1-2. Calibration ComponentsNotes:*Calibration gas bottles can
HIGHLIGHTS OF CHANGESEffective January 2007 Rev 4.3Effective July 2008 Rev 4.4Page SummaryGeneral Revised reference air specifications to read 0.25 l/
Hazardous Area Oxymitter 40001-20Instruction ManualIM-106-340C, Rev. 4.4July 2008
Instruction Manual IM-106-340C, Rev. 4.4July 2008Hazardous Area Oxymitter 4000http://www.raihome.comSection 2 InstallationMechanical Installation .
Hazardous Area Oxymitter 40002-2Instruction ManualIM-106-340C, Rev. 4.4July 2008MECHANICAL INSTALLATIONSelecting Location1. The location of the Hazard
Instruction Manual IM-106-340C, Rev. 4.4July 20082-3Hazardous Area Oxymitter 4000Figure 2-1. Hazardous Area Oxymitter 4000 Probe InstallationCoverRem
Hazardous Area Oxymitter 40002-4Instruction ManualIM-106-340C, Rev. 4.4July 2008Figure 2-2. Hazardous Area Oxymitter 4000 Remote Electronics Installat
Instruction Manual IM-106-340C, Rev. 4.4July 20082-5Hazardous Area Oxymitter 4000Figure 2-3. Hazardous Area Oxymitter 4000 Probe with Abrasive ShieldP
Hazardous Area Oxymitter 40002-6Instruction ManualIM-106-340C, Rev. 4.4July 2008Figure 2-4. Hazardous Area Oxymitter 4000 Mounting Plate Dimensions22.
Instruction Manual IM-106-340C, Rev. 4.4July 20082-7Hazardous Area Oxymitter 4000Figure 2-5. Hazardous Area Oxymitter 4000 Mounting Plate Installation
Hazardous Area Oxymitter 40002-8Instruction ManualIM-106-340C, Rev. 4.4July 20084. If using the optional ceramic diffusion element, the vee deflector
Instruction Manual IM-106-340C, Rev. 4.4July 20082-9Hazardous Area Oxymitter 4000Figure 2-6. Orienting the Optional Vee DeflectorFigure 2-7. Installat
Hazardous Area Oxymitter 40002-10Instruction ManualIM-106-340C, Rev. 4.4July 2008Remote Electronics Installation For a Hazardous Area Oxymitter 4000 w
Instruction Manual IM-106-340C, Rev. 4.4July 20082-11Hazardous Area Oxymitter 4000NOTETo maintain CE compliance, ensure a good connection exists betwe
Hazardous Area Oxymitter 40002-12Instruction ManualIM-106-340C, Rev. 4.4July 2008Figure 2-9. Electrical Installation - Hazardous Area Oxymitter 4000 w
Instruction Manual IM-106-340C, Rev. 4.4July 20082-13Hazardous Area Oxymitter 4000ELECTRICAL INSTALLATION WITH REMOTE ELECTRONICSAll wiring must confo
Hazardous Area Oxymitter 40002-14Instruction ManualIM-106-340C, Rev. 4.4July 2008Connect Line Voltage1. Remove screw (18, Figure 9-4), cover lock (19)
Instruction Manual IM-106-340C, Rev. 4.4July 20082-15Hazardous Area Oxymitter 4000Figure 2-10. Electrical Installation - Hazardous Area Oxymitter 4000
Hazardous Area Oxymitter 40002-16Instruction ManualIM-106-340C, Rev. 4.4July 2008Figure 2-10. Electrical Installation - Hazardous Area Oxymitter 4000
Instruction Manual IM-106-340C, Rev. 4.4July 20082-17Hazardous Area Oxymitter 4000Calibration Handshake Logic I/O1. The output signal can be used to t
Hazardous Area Oxymitter 40002-18Instruction ManualIM-106-340C, Rev. 4.4July 2008PNEUMATIC INSTALLATIONReference Air PackageAfter the Hazardous Area O
Instruction Manual IM-106-340C, Rev. 4.4July 20082-19Hazardous Area Oxymitter 4000Figure 2-12. Hazardous Area Oxymitter 4000 Gas ConnectionsCalibratio
Instruction ManualIM-106-340C, Rev. 4.4July 2008TOC-1Hazardous Area Oxymitter 4000Table of ContentsEssential Instructions. . . . . . . . . . . . . . .
Hazardous Area Oxymitter 40002-20Instruction ManualIM-106-340C, Rev. 4.4July 2008
Instruction Manual IM-106-340C, Rev. 4.4July 2008Hazardous Area Oxymitter 4000http://www.raihome.comSection 3 Configuration of Hazardous Area Oxymitt
Hazardous Area Oxymitter 40003-2Instruction ManualIM-106-340C, Rev. 4.4July 2008Figure 3-1. Electronics Housing Terminals and Membrane Keypad Hazardou
Instruction Manual IM-106-340C, Rev. 4.4July 20083-3Hazardous Area Oxymitter 4000SW2 SettingThe factory sets this switch as follows:1. Position 1 is H
Hazardous Area Oxymitter 40003-4Instruction ManualIM-106-340C, Rev. 4.4July 2008Figure 3-2. Defaults - Hazardous Area Oxymitter 4000 with Membrane Key
Instruction Manual IM-106-340C, Rev. 4.4July 20083-5Hazardous Area Oxymitter 4000LOGIC I/OThis two-terminal logic contact can be configured either as
Hazardous Area Oxymitter 40003-6Instruction ManualIM-106-340C, Rev. 4.4July 2008Table 3-1. Logic I/O Configuration (as set at HART/AMS or LOI)*The de
Instruction Manual IM-106-340C, Rev. 4.4July 20083-7Hazardous Area Oxymitter 40002. IN CALIBRATION. One contact per probe provides notification to the
Hazardous Area Oxymitter 40003-8Instruction ManualIM-106-340C, Rev. 4.4July 2008
Instruction Manual IM-106-340C, Rev. 4.4July 2008Hazardous Area Oxymitter 4000http://www.raihome.comSection 4 Configuration of Hazardous Area Oxymitt
Instruction ManualIM-106-340C, Rev. 4.4July 2008Hazardous Area Oxymitter 4000TOC-2SECTION 4Configuration of Hazardous Area Oxymitter 4000 with LOIVeri
Hazardous Area Oxymitter 40004-2Instruction ManualIM-106-340C, Rev. 4.4July 2008Figure 4-1. Electronics Housing Terminals with LOI Hazardous Area Oxym
Instruction Manual IM-106-340C, Rev. 4.4July 20084-3Hazardous Area Oxymitter 4000SW2The factory sets this switch as follows:1. Position 1 is HART/LOCA
Hazardous Area Oxymitter 40004-4Instruction ManualIM-106-340C, Rev. 4.4July 2008Figure 4-2. Defaults - Hazardous Area Oxymitter 4000 with LOI LOGIC I/
Instruction Manual IM-106-340C, Rev. 4.4July 20084-5Hazardous Area Oxymitter 4000AlarmWhen configured as an alarm, this signal alerts you to an out-of
Hazardous Area Oxymitter 40004-6Instruction ManualIM-106-340C, Rev. 4.4July 2008Recommended Configuration4-20 mA Signal Upon Critical AlarmEmerson Pro
Instruction Manual IM-106-340C, Rev. 4.4July 2008Hazardous Area Oxymitter 4000http://www.raihome.comSection 5 Startup and Operation of Hazardous Area
Hazardous Area Oxymitter 40005-2Instruction ManualIM-106-340C, Rev. 4.4July 2008ErrorIf there is an error condition at startup, one of the diagnostics
Instruction Manual IM-106-340C, Rev. 4.4July 20085-3Hazardous Area Oxymitter 4000Figure 5-2. Calibration Keys Figure 5-3. Normal Operation DIAGNOSTICA
Hazardous Area Oxymitter 40005-4Instruction ManualIM-106-340C, Rev. 4.4July 2008TEST POINTSTest points 1 through 6 allow you to monitor with a multime
Instruction Manual IM-106-340C, Rev. 4.4July 2008Hazardous Area Oxymitter 4000http://www.raihome.comSection 6 Startup and Operation of Hazardous Area
Instruction ManualIM-106-340C, Rev. 4.4July 2008TOC-3Hazardous Area Oxymitter 4000Replacement of Entire Electronics (with Housing) . . . . . . . . . .
Hazardous Area Oxymitter 40006-2Instruction ManualIM-106-340C, Rev. 4.4July 2008Figure 6-1. Startup Display Figure 6-2. Normal Display 38860039O2:Ø
Instruction Manual IM-106-340C, Rev. 4.4July 20086-3Hazardous Area Oxymitter 4000Figure 6-3. LOI FeaturesSTART UP OXYMITTER 4000 CALIBRATIONRefer to S
Hazardous Area Oxymitter 40006-4Instruction ManualIM-106-340C, Rev. 4.4July 2008In order to unlock the display, input a "Z" pattern. First,
Instruction Manual IM-106-340C, Rev. 4.4July 20086-5Hazardous Area Oxymitter 4000Figure 6-4. Local Operator Interface Menu Tree (Sheet 1 of 2) Tempera
Hazardous Area Oxymitter 40006-6Instruction ManualIM-106-340C, Rev. 4.4July 2008Figure 6-4. Local Operator Interface Menu Tree (Sheet 2 of 2) HAZARDOU
Instruction Manual IM-106-340C, Rev. 4.4July 20086-7Hazardous Area Oxymitter 4000NOTEEmerson Process Management recommends 0.4% O2 and 8% O2 for calib
Hazardous Area Oxymitter 40006-8Instruction ManualIM-106-340C, Rev. 4.4July 2008SYSTEM/ParametersO2 Slope - O2 slope is data regarding the strength of
Instruction Manual IM-106-340C, Rev. 4.4July 20086-9Hazardous Area Oxymitter 4000TemperaturesO2 Temp - Indicates the thermocouple temperature at the s
Hazardous Area Oxymitter 40006-10Instruction ManualIM-106-340C, Rev. 4.4July 2008OXYMITTER 4000 TEST POINTSRefer to Figure 6-6. System test points ar
Instruction Manual IM-106-340C, Rev. 4.4July 2008Hazardous Area Oxymitter 4000http://www.raihome.comSection 7 HART/AMSOverview . . . . . . . . . . .
Instruction ManualIM-106-340C, Rev. 4.4July 2008Hazardous Area Oxymitter 4000TOC-4
Hazardous Area Oxymitter 40007-2Instruction ManualIM-106-340C, Rev. 4.4July 2008HART COMMUNICATOR SIGNAL LINE CONNECTIONSThe HART Communicator can con
Instruction Manual IM-106-340C, Rev. 4.4July 20087-3Hazardous Area Oxymitter 4000Figure 7-1. Signal Line Connections, ≥ 250 Ohms Load Resistance LOOP
Hazardous Area Oxymitter 40007-4Instruction ManualIM-106-340C, Rev. 4.4July 2008Figure 7-2. Signal Line Connections, < 250 Ohms Load Resistance HAR
Instruction Manual IM-106-340C, Rev. 4.4July 20087-5Hazardous Area Oxymitter 4000The opening menu displayed on the LCD is different for on-line and of
Hazardous Area Oxymitter 40007-6Instruction ManualIM-106-340C, Rev. 4.4July 2008Figure 7-3. HART/AMS Menu Tree (Sheet 1 of 3) O2valueO2celltempCJ
Instruction Manual IM-106-340C, Rev. 4.4July 20087-7Hazardous Area Oxymitter 4000Figure 7-3. HART/AMS Menu Tree (Sheet 2 of 3) DEVICESETUPPVPV AOPV
Hazardous Area Oxymitter 40007-8Instruction ManualIM-106-340C, Rev. 4.4July 2008Figure 7-3. HART/AMS Menu Tree (Sheet 3 of 3) DEVICESETUPPVPV AOPV
Instruction Manual IM-106-340C, Rev. 4.4July 20087-9Hazardous Area Oxymitter 4000HART COMMUNICATOR O2 CAL METHODUse the following procedure to perfo
Hazardous Area Oxymitter 40007-10Instruction ManualIM-106-340C, Rev. 4.4July 20089. From the PERFORM O2 CAL screen, view menu item 3, CALSTATE, to mon
Instruction Manual IM-106-340C, Rev. 4.4July 20087-11Hazardous Area Oxymitter 4000D/A TRIM PROCEDUREThe D/A trim procedure is used to calibrate the 4-
Instruction Manual IM-106-340C, Rev. 4.4July 2008Hazardous Area Oxymitter 4000http://www.raihome.comOxymitter Oxygen TransmittersREAD THIS PAGE BEFORE
Hazardous Area Oxymitter 40007-12Instruction ManualIM-106-340C, Rev. 4.4July 2008
Instruction Manual IM-106-340C, Rev. 4.4July 2008Hazardous Area Oxymitter 4000http://www.raihome.comSection 8 TroubleshootingOverview . . . . . . .
Hazardous Area Oxymitter 40008-2Instruction ManualIM-106-340C, Rev. 4.4July 2008Figure 8-1. O2 Sensor mV Reading vs. % O2 at 736°C (Reference Air, 20.
Instruction Manual IM-106-340C, Rev. 4.4July 20088-3Hazardous Area Oxymitter 4000GENERALThe troubleshooting section describes how to identify and isol
Hazardous Area Oxymitter 40008-4Instruction ManualIM-106-340C, Rev. 4.4July 2008Figure 8-2. Diagnostic LEDs ALARM CONTACTSIf autocalibration is not u
Instruction Manual IM-106-340C, Rev. 4.4July 20088-5Hazardous Area Oxymitter 4000SPS 4001B and IMPS 4000, 1-4 probes• One contact closure per probe fr
Hazardous Area Oxymitter 40008-6Instruction ManualIM-106-340C, Rev. 4.4July 2008Table 8-1. Diagnostic/Unit Alarm Fault Definitions - Membrane Keypad
Instruction Manual IM-106-340C, Rev. 4.4July 20088-7Hazardous Area Oxymitter 4000Figure 8-3. Fault 1, Open ThermocoupleFault 1, Open ThermocoupleFigur
8-8Instruction ManualIM-106-340C, Rev. 4.4July 2008Hazardous Area Oxymitter 4000Figure 8-4. Fault 2, Shorted ThermocoupleFault 2, Shorted Thermocouple
Instruction Manual IM-106-340C, Rev. 4.4July 20088-9Hazardous Area Oxymitter 4000Figure 8-5. Fault 3, Reversed ThermocoupleFault 3, Reversed Thermocou
Commentaires sur ces manuels